CAN: Procurement of Repl SO2 NDIR Gas Analysers

Award

Value

£52,878

Suppliers

Classifications

  • Laboratory, optical and precision equipments (excl. glasses)

Tags

  • award

Submission Deadline

1 year ago

Published

9 months ago

Description

The National Physical Laboratory (NPL) requirement is for replacement NDIR and paramagnetic sensor analysers, with the following capabilities:
1.	Capable of measuring the following gases below their respective amount fractions:
a.	SO2 - 1 %
b.	O2 - 25 %
c.	CO - 5 %
d.	CO2 - 10 %
2.	The analyser shall be CE/UKCA certificated.
3.	There must be a datasheet, manual and user instructions
4.	Analog output (4-20 mA or 0-10 Vdc) is required to allow connectivity to exiting automatic analysers and LabView integration. 
5.	Flowrate: ≤ 1 L min-1
6.	Linearity error: ≤ 1.5 % of span 
7.	Repeatability: ≤ 1 % of span
8.	The analyser must be suitable for 19" rack installation 
9.	Permitted pressure: 800 - 1200 kPa
10.	Detection limit (4 σ): ≤ 1 % of span

Documents

Premium

Bypass the hassle of outdated portals. Get all the information you need right here, right now.

  • Contract Agreement

    The official contract terms, conditions, and scopes of work.

    Download
  • Award Notice

    Details on the tender award and selected suppliers.

    Download

Similar Contracts

AI Bid Assistant

Our AI-powered tool to help you create winning bids is coming soon!

View Contract Source Save Contract

Timeline complete

Publish
Bid
Evaluate
Award
Complete